Enciclopedia della Tecnologia di Rilevamento a Reticolo di Bragg su Fibra (FBG) - FAQ FBG - FAQ Sensore FBG - FAQ Demodulatore FBG - Analisi dei Problemi Comuni di Applicazione FBG
Coefficiente di temperatura del sensore a reticolo di Bragg in fibra a femtosecondi (FBG) - Funzione di calibrazione di fabbrica per termometro FBG da 800°C, con un intervallo da -16 a 806°C e da 524 a 1000°C
The OFSCN® 800°C Capillary Seamless Steel Tube Fiber Bragg Grating temperature sensor (FBG thermometer) , manufactured by DCYS, undergoes temperature calibration before leaving the factory. Along with the product, basic temperature coefficients (also known as calibration coefficients or calibration functions) of the fiber optic grating sensor are provided. The following is a brief description of the temperature calibration of the product at the time of shipment:
Indice:
| I. Struttura del Prodotto del Sensore di Temperatura a Reticolo di Bragg su Fibra Femtosecondo Capillary® da 800°C |
I. Product Structure of the OFSCN® 800°C Capillary Femtosecond Fiber Bragg Grating Temperature Sensor
The OFSCN® 800°C Capillary Seamless Steel Tube Fiber Bragg Grating Temperature Sensor (FBG thermometer) adopts a fully metallic packaging method to protect the FBG. The standard product has a diameter of 2.0 millimeters and consists of a three-layer seamless steel tube nested structure. It encapsulates a high-temperature-resistant femtosecond fiber bragg grating (Femtosecond FBG). The overall structure and product are shown in Figura 1, 2, and 3 below:
![]() |
|
|
Figura 1 |
|
![]() |
![]() |
|
Figure 2 |
Figura 3 |
I connettori e le terminazioni sono mostrati nelle Figure 4 e 5 di seguito:
![]() |
![]() |
|
Figure 4 |
Figure 5 |
Due to the unique stainless steel seamless tube patented packaging technology used in the OFSCN® 800°C Capillary Seamless Steel Tube Fiber Bragg Grating Temperature Sensor (FBG thermometer), as well as our special process treatment, it possesses high sensitivity, fast thermal conduction, good accuracy, small size, light weight, high tensile strength, high pressure resistance, high temperature resistance, water e moisture resistance, non-combustibility, e corrosion resistance.
II. Factory Test Report of the OFSCN® 800°C Capillary® Seamless Steel Pipe Femtosecond Fiber Bragg Grating Temperature Senso - Temperature Coefficient Calibration Graph
High accuracy is the lifeline of DCYS's products, a feature that is verified through the test reports of each sensor produced, as shown in the following figure:
![]() |
|
Figura 6 |
Figura 6 displays the actual measurement graph of the temperature coefficient (calibration coefficient, calibration function) of a sold OFSCN® 800°C Capillary Seamless Steel Tube Fiber Bragg Grating Temperature Sensor (FBG thermometer). This curve shows the correspondence between the sensor's wavelength variation and temperature change. The test range is from -16 to 806°C, with temperature calibration intervals of approximately 30°C (not strictly enforced).
![]() |
|
Figura 7 |
The OFSCN® 800°C Capillary Seamless Steel Tube Fiber Bragg Grating Temperature Sensor (FBG thermometer) can be temporarily used at higher temperatures (duration of use not tested, repeatability to be determined). Figura 7 presents the actual measurement graph of the temperature coefficient (calibration coefficient, calibration function) of a specific OFSCN® 800°C Capillary Seamless Steel Tube FBG Temperature Sensor (FBG thermometer). This curve shows the correspondence between the sensor's wavelength variation and temperature change. The main test temperature range is from 524 to 1001°C (initial temperature of 53°C), with temperature calibration intervals of approximately 40°C (not strictly enforced).
Contrariamente alla legenda, nel rapporto ufficiale di fabbrica, verranno forniti valori specifici di lunghezza d'onda e temperatura. Per misurazioni generali della temperatura, questo coefficiente di calibrazione può essere usato direttamente.
Each OFSCN® 800°C Capillary Seamless Steel Tube Fiber Bragg Grating Temperature Sensor has a unique temperature coefficient (calibration coefficient, calibration function). This coefficient (function) is tested e determined by the Quality Inspection Department of DCYS e recorded on the company's official website for customers to check at any time. Based on our experience, the R-square value of the binomial function is generally above 0.9995.
Our philosophy is: 'OFSCN®, make optical fiber stronger!'
DCYS is a professional 800°C FBG temperature sensor manufacturer, providing 800°C FBG temperature sensor calibration method, calibration experiment, test report, temperature coefficient, calibration function and Altro information; 800°C fiber bragg grating temperature sensor brand is 'OFSCN'.