In Primo Piano

Coefficiente di temperatura del sensore a reticolo di Bragg in fibra a femtosecondi (FBG) - Funzione di calibrazione di fabbrica per termometro FBG da 800°C, con un intervallo da -16 a 806°C e da 524 a 1000°C

The OFSCN® 800°C Capillary Seamless Steel Tube Fiber Bragg Grating temperature sensor (FBG thermometer) , manufactured by DCYS, undergoes temperature calibration before leaving the factory. Along with the product, basic temperature coefficients (also known as calibration coefficients or calibration functions) of the fiber optic grating sensor are provided. The following is a brief description of the temperature calibration of the product at the time of shipment:

Indice:

I. Struttura del Prodotto del Sensore di Temperatura a Reticolo di Bragg su Fibra Femtosecondo Capillary® da 800°C

II. Rapporto di Calibrazione di Fabbrica del Sensore di Temperatura a Reticolo di Bragg su Fibra Femtosecondo - Coefficiente di Temperatura a -1000°C

I. Product Structure of the OFSCN® 800°C Capillary Femtosecond Fiber Bragg Grating Temperature Sensor

The OFSCN® 800°C Capillary Seamless Steel Tube Fiber Bragg Grating Temperature Sensor (FBG thermometer) adopts a fully metallic packaging method to protect the FBG. The standard product has a diameter of 2.0 millimeters and consists of a three-layer seamless steel tube nested structure. It encapsulates a high-temperature-resistant femtosecond fiber bragg grating (Femtosecond FBG). The overall structure and product are shown in Figura 1, 2, and 3 below:

 

Diagramma fisico del sensore di temperatura/deformazione/stress FBG per tubo capillare in acciaio senza saldatura OFSCN® (parte di esso)

Figura 1

Diagramma fisico del sensore di temperatura FBG con tubo capillare in acciaio senza saldatura OFSCN® Struttura principale del sensore FBG con tubo capillare in acciaio senza saldatura OFSCN®

Figure  2

Figura 3


I connettori e le terminazioni sono mostrati nelle Figure 4 e 5 di seguito:

Connettore in fibra del sensore FBG con tubo capillare in acciaio senza saldatura OFSCN® Estremità finale del sensore FBG con tubo capillare in acciaio senza saldatura OFSCN®

Figure 4

Figure 5

Due to the unique stainless steel seamless tube patented packaging technology used in the OFSCN® 800°C Capillary Seamless Steel Tube Fiber Bragg Grating Temperature Sensor (FBG thermometer), as well as our special process treatment, it possesses high sensitivity, fast thermal conduction, good accuracy, small size, light weight, high tensile strength, high pressure resistance, high temperature resistance, water e moisture resistance, non-combustibility, e corrosion resistance.

 

II. Factory Test Report of the OFSCN® 800°C Capillary® Seamless Steel Pipe Femtosecond Fiber Bragg Grating Temperature Senso - Temperature Coefficient Calibration Graph 

High accuracy is the lifeline of DCYS's products, a feature that is verified through the test reports of each sensor produced, as shown in the following figure:

Diagramma misurato della lunghezza d'onda e della temperatura per il sensore di temperatura FBG con tubo capillare in acciaio senza saldatura OFSCN® 800°C

Figura 6

Figura 6 displays the actual measurement graph of the temperature coefficient (calibration coefficient, calibration function) of a sold OFSCN® 800°C Capillary Seamless Steel Tube Fiber Bragg Grating Temperature Sensor (FBG thermometer). This curve shows the correspondence between the sensor's wavelength variation and temperature change. The test range is from -16 to 806°C, with temperature calibration intervals of approximately 30°C (not strictly enforced).

Diagramma di calibrazione della lunghezza d'onda e della temperatura del sensore di temperatura FBG con tubo capillare in acciaio senza saldatura OFSCN® 800°C

Figura 7

The OFSCN® 800°C Capillary Seamless Steel Tube Fiber Bragg Grating Temperature Sensor (FBG thermometer) can be temporarily used at higher temperatures (duration of use not tested, repeatability to be determined). Figura 7 presents the actual measurement graph of the temperature coefficient (calibration coefficient, calibration function) of a specific OFSCN® 800°C Capillary Seamless Steel Tube FBG Temperature Sensor (FBG thermometer). This curve shows the correspondence between the sensor's wavelength variation and temperature change. The main test temperature range is from 524 to 1001°C (initial temperature of 53°C), with temperature calibration intervals of approximately 40°C (not strictly enforced).

Contrariamente alla legenda, nel rapporto ufficiale di fabbrica, verranno forniti valori specifici di lunghezza d'onda e temperatura. Per misurazioni generali della temperatura, questo coefficiente di calibrazione può essere usato direttamente.

Each OFSCN® 800°C Capillary Seamless Steel Tube Fiber Bragg Grating Temperature Sensor has a unique temperature coefficient (calibration coefficient, calibration function). This coefficient (function) is tested e determined by the Quality Inspection Department of DCYS e recorded on the company's official website for customers to check at any time. Based on our experience, the R-square value of the binomial function is generally above 0.9995.

Our philosophy is: 'OFSCN®, make optical fiber stronger!'

  • DCYS is a professional 800°C FBG temperature sensor manufacturer, providing 800°C FBG temperature sensor calibration method, calibration experiment, test report, temperature coefficient, calibration function and Altro information; 800°C fiber bragg grating temperature sensor brand is 'OFSCN'.

💬 OFSCN Official Knowledge Base