Koeffizient des Femtosekunden-Faser-Bragg-Gitter (FBG) Temperatursensors – Werkskalibrierfunktion für 500°C FBG Thermometer, Bereich von -4.6℃ bis 450℃

OFSCN® 500°C Capillary Seamless Steel Tube Fiber Bragg Grating Temperature Sensors (FBG Thermometers) produced by DCYS undergo calibration before leaving the factory, und the basic temperature coefficients (also known as calibration coefficients or calibration functions) for the fiber bragg grating sensors are provided along with the product. Here is a brief explanation of the calibration for this product:

Inhaltsverzeichnis:

1. Produktstruktur des Femtosekunden-Faser-Bragg-Gitter-Temperatursensors

2. Werksprüfbericht für Femtosekunden-Faser-Bragg-Gitter-Temperatursensor - Kalibrierungstemperaturkoeffizienten

1.Product Structure of OFSCN® 500°C Capillary® Seamless Steel Tube Femtosecond Fiber Bragg Grating Temperature Sensor:

OFSCN® 500°C Capillary Seamless Steel Tube Fiber Bragg Grating Temperature Sensors (FBG Thermometers) are protected by a full-metal encapsulation to safeguard the fiber bragg grating (FBG). The standard product has a diameter of 2.0 millimeters and features a three-layer metal pipe nested structure. It encapsulates a high-temperature-resistant femtosecond fiber bragg grating (Femtosecond FBG). The overall structure and the product itself are shown in Abbildung 1, 2, and 3:

 

Physikalisches Diagramm des OFSCN® Kapillarnahtlosen Stahlrohr-FBG-Temperatur-/Dehnungs-/Spannungssensors (Teil davon)

Abbildung 1

Physikalisches Diagramm des OFSCN® FBG-Temperatursensors mit nahtlosem Kapillarstahlrohr Hauptstruktur des OFSCN® kapillaren nahtlosen Stahlrohr-FBG-Sensors
Abbildung 2

Abbildung 3


Die Steckverbinder und Endstücke sind in den Abbildungen 4 und 5 dargestellt:

Faserverbinder des OFSCN® Kapillarnahtlosen Stahlrohr-FBG-Sensors Endstück des OFSCN® Kapillarnahtlosen Stahlrohr-FBG-Sensors

Abbildung 4

Abbildung 5

2. Factory Test Report for OFSCN® 500°C Capillary Seamless Steel Tube Femtosecond Fiber Bragg Grating Temperature Sensor - Calibration Coefficient Graph:

High accuracy is the lifeline of DCYS's products, und this characteristic is verified through the test reports of each sensor we produce, as shown in the following figure:

 

Gemessenes Diagramm der Wellenlänge und Temperatur für den FBG-Temperatursensor OFSCN® 350 °C mit nahtlosem Kapillarstahlrohr
Abbildung 6

 

Abbildung 6 shows the measured graph of the calibration coefficient (calibration function) of an actual sold OFSCN® 500°C Capillary Seamless Steel Tube Fiber Bragg Grating Temperature Sensor (FBG Thermometer). This curve displays the corresponding relationship between the sensor's wavelength variation and temperature change. As this sensor can be used at higher temperatures, we provide calibration for temperatures ranging from room temperature to 450°C, with calibration temperature intervals of approximately 30°C (not strictly enforced).

Im Gegensatz zur Legende stellen wir im formellen Werksbericht auch spezifische Wellenlängen- und Temperaturwerte zur Verfügung. Für allgemeine Temperaturmessungen kann dieser Kalibrierungskoeffizient direkt verwendet werden.

Each OFSCN® 500°C Capillary Seamless Steel Tube Fiber Bragg Grating Temperature Sensor has a unique calibration coefficient (calibration function), which is tested und determined by the Quality Inspection Department of DCYS. It is recorded on the company's official website for customers to query at any time. Based on our experience, the R-square value of the binomial function is generally above 0.9995.

Our philosophy is: 'OFSCN®, make optical fiber stronger!'

  • DCYS is a professional 500°C FBG temperature sensor manufacturer, providing 500°C FBG temperature sensor calibration method, calibration experiment, test report, temperature coefficient, calibration function and Sonstiges information; 500°C fiber bragg grating temperature sensor brand is 'OFSCN'.