ファイバーブラッググレーティングセンシング技術百科事典 - FBG FAQ - FBGセンサー FAQ - FBG復調器 FAQ - FBG一般的なアプリケーション問題分析
光ファイバブラッググレーティング(FBG)温度センサーの係数 - 100°C FBG温度計の校正機能
OFSCN® 100-degree Celsius capillary seamless steel tube fiber bragg grating (FBG) temperature sensors (FBG thermometers) produced by DCYS undergo calibration during the manufacturing process. They are accompanied by the basic temperature coefficient (also known as calibration coefficient or calibration function) of the fiber bragg grating sensor. The following is a brief description of the calibration of this product:
目次:
| 1. 100°C ファイバーブラッググレーティング温度センサーの製品構造 |
1. Product Structure of OFSCN® 100°C Capillary Seamless Steel Tube Fiber Bragg Grating Temperature Sensor:
OFSCN® 100°C Capillary Seamless Steel Tube Fiber Bragg Grating Temperature Sensors (FBG Thermometers) are protected by a full-metal encapsulation to safeguard the fiber bragg grating (FBG). The standard product has a diameter of 0.9 millimeters, and the overall structure and the product itself are shown in 図1, 2, and 3:
![]() |
|
|
図1 |
|
![]() |
![]() |
|
図2 |
図3 |
コネクタとテールエンドは図4と図5に示されています:
![]() |
![]() |
|
図4 |
図5 |
Due to the unique stainless steel seamless tube patent packaging technology および our special processing techniques applied to the OFSCN® 100-degree Celsius capillary seamless steel tube fiber bragg grating temperature sensor (FBG thermometer), it possesses the following features: high sensitivity, fast thermal conduction, good accuracy, small size, light weight, high tensile および compressive strength, high temperature resistance, waterproof および moisture-proof properties, non-flammable, および corrosion resistance.
2. Factory Test Report for OFSCN® 100°C Capillay Seamless Steel Tube Fiber Bragg Grating Temperature Sensor - Calibration Coefficient Graph:
High accuracy is the lifeline of DCYS's products, および this characteristic is verified through the test reports of each sensor we produce, as shown in the following figure:
![]() |
|
図6 |
図6 above shows the measured chart of the calibration coefficient (calibration function) of an actual sold OFSCN® 100-degree Celsius capillary seamless steel tube fiber bragg grating temperature sensor. The curve illustrates the correlation between the wavelength variation of the sensor and temperature change, with a test range from room temperature to 100 degrees Celsius. Since this sensor can be used up to a maximum of 120 degrees Celsius, if requested by the customer, we can also provide calibration for temperatures ranging from room temperature to 120 degrees Celsius, with a calibration temperature interval of approximately 10 degrees Celsius.
凡例とは異なり、公式の工場レポートでは特定の波長値と温度値も提供されます。一般的な温度測定要件については、この校正係数を直接使用できます。
Each OFSCN® 100-degree Celsius capillary seamless steel tube fiber bragg grating temperature sensor (FBG thermometer) has a unique calibration coefficient (calibration function), which is tested および recorded by the Quality Inspection Department of DCYS, および can be accessed by customers at any time on the company's official website. In our experience, the R-squared value of the polynomial equation for the calibration function can reach 0.9995 or higher.
Our philosophy is: 'OFSCN®, Make Optical Fiber Stronger!'
DCYS is a professional 100℃ FBG温度センサー manufacturer, providing 100℃ FBG温度センサー calibration method, calibration experiment, test report, temperature coefficient, calibration function and other information; 100°C fiber bragg grating temperature sensor brand is 'OFSCN'.